Research Article

Evaluation of Three Human-Use Glucometers for Blood Glucose Measurement in Dogs

Table 5

Percentage and absolute values of blood glucose concentration obtained using the three glucometers, according to ISO 15197 : 2013 standards in dogs.

Established valuesFreedom LiteOptiumNeoOnCall Plus
nTotal%nTotal%NTotal%

<100 mg/dL: Difference of+/- 15 mg/dL85016485096495098
≥100 mg/dL: Difference of +/-15%0650.0586589.2426564.6
Total within acceptable limits81157.010611592.29111579.1