Research Article

Evaluation of Three Human-Use Glucometers for Blood Glucose Measurement in Dogs

Table 4

Percentage and absolute values of blood glucose concentration obtained using the three glucometers, according to the ISO 15197 : 2003 standards in dogs.

Established valuesFreedom LiteOptiumNeoOnCall Plus
nTotal%nTotal%nTotal%

<75 mg/dL: Difference of +/-15 mg/dL49449910099100
≥75 mg/dL: Difference of +/- 20%101069.410510699.18910684.0