Journals
Publish with us
Publishing partnerships
About us
Blog
Veterinary Medicine International
Journal overview
For authors
For reviewers
For editors
Table of Contents
Special Issues
Veterinary Medicine International
/
2022
/
Article
/
Tab 4
/
Research Article
Evaluation of Three Human-Use Glucometers for Blood Glucose Measurement in Dogs
Table 4
Percentage and absolute values of blood glucose concentration obtained using the three glucometers, according to the ISO 15197 : 2003 standards in dogs.
Established values
Freedom Lite
OptiumNeo
OnCall Plus
n
Total
%
n
Total
%
n
Total
%
<75 mg/dL: Difference of +/-15 mg/dL
4
9
44
9
9
100
9
9
100
≥75 mg/dL: Difference of +/- 20%
10
106
9.4
105
106
99.1
89
106
84.0
Total with inacceptable limits
14
115
12.2
114
115
99.1
98
115
85.2