Validation of Two Activity Monitors in Slow and Fast Walking Hospitalized Patients

Table 3

Agreement between steps and walking measured by direct observations and by ActivPAL3™ and SENS motion® used as both single- and dual-sensor systems.

StepsWalking time

Assessment method
All gait speeds
Slow walkers
Fast walkers
All gait speeds
Slow walkers
Fast walkers
#
Direct observation7738381138656397#3758#2446#
ActivPAL, single sensor5096187232003863#1778#2039#
SENS, single sensor5800231133115506#3175#2183#
ActivPAL, dual sensor5096187232003795#1725#2040#
SENS, dual sensor5800243533115506#3175#2182#
Percentage agreement between direct observation and ActivPAL/SENS
ActivPAL (single sensor), median % (IQR)71.9 (39.5-80.4)45.9 (10.7-74.7)82.3 (75.1-90.6)73.1 (29.3-87.6)59.6 (9.4-82.4)83.9 (73.1-92.1)
SENS (single sensor), median % (IQR)78.9 (40.3-87.1)53.33 (38.8-81.4)83.6 (78.4-91.4)88.9 (84.6-94.2)89.4 (81.3-92.6)89.4 (86.3-95.7)
ActivPAL (dual sensor), median % (IQR)71.9 (39.5-80.4)45.9 (10.7-74.7)82.3 (75.1-90.6)73.1 (29.3-87.6)59.2 (9.3-82.4)83.0 (73.9-95.1)
SENS (dual sensor), median % (IQR)78.9 (40.3-87.1)53.33 (38.8-81.4)83.6 (78.4-91.4)88.9 (84.6-94.2)89.4 (81.3-92.6)89.4 (86.3-95.7)

Number of steps; #walking time in sec; slow walkers: <0.67 m/s; fast walkers: ≥0.67 m/s. ActivPAL: ActivPAL3™ monitor; SENS: SENS motion®.