A Comparative Analysis of Methods for Evaluation of ECG Signal Quality after Compression

Table 2

Boundary values of avL for classification of compressed ECG signals (leads I and V1) into 3 quality groups (a: perfect quality, b: good quality, and c: not evaluable ECG). The values come from two experts (1 and 2). The overall avL boundaries can be set as strict-max case, mild-min case, and medium-median case.

ā€‰lead V1: bound a/blead V1: bound b/clead I: bound a/blead I: bound b/c
expert 1expert 2expert 1expert 2expert 1expert 2expert 1expert 2

min avL0.200.200.100.100.150.200.100.10
max avL0.700.800.200.250.650.650.200.20
median avL0.400.350.100.150.300.300.100.15