Automation of the Peak Fitting Method in Bone FTIR Microspectroscopy Spectrum Analysis: Human and Mice Bone Study

Table 2

Differences of the manual peak fitting analysis intraoperator, interoperator, maximum absolute error, and variation of results using the automatic method for each parameter, based on the analysis of 5 bone spectra. Operator 1 is an experienced operator, and operator 2 is an inexperienced operator to whom has been given the instructions to analyse the spectra.

Crystallinity IndexMineral MaturityCollagen Maturity

RMS of coefficient of variation (operator 1) (%)0.71.40.7
RMS of coefficient of variation (operator 2) (%)1.70.41.0
RMS of coefficient of variation (interoperator) (%)2.01.20.7
Maximum absolute error0.0013 cm0.070.11
Variation of results using automatic methodNo variationNo variationNo variation

RMS = root mean square.