Abstract

In this work, carbon quantum dots (CQDs) were synthesized by microwave irradiation and were electropolymerized on glassy carbon electrode (GCE) to establish an electrochemical sensor for the selective detection of ascorbic acid (AA). Electrochemical behaviors of the prepared sensor were investigated by cyclic voltammetry (CV), differential pulse voltammetry (DPV), and electrochemical impedance spectroscopy (EIS). Herein, two wide linear responses were obtained in ranges of 0.01-3 mM and 4-12 mM with a low detection limit of 10 μM to AA. High sensitivities (44.13 μA-1μM-1 cm-2, 9.66 μA-1μM-1 cm-2, respectively) corresponding to the linear ranges were also achieved. In addition, the electrochemical sensor exhibited good selectivity and robust anti-interference ability toward AA in the presence of dopamine (DA) and uric acid (UA). These results showed that this sensor can be used as a promising tool to detect AA in real complex systems.

1. Introduction

Ascorbic acid (vitamin C, AA) plays a crucial role as an essential nutrient and antioxidant in the human body, which exists only in food and drugs and cannot be synthesized by human beings themselves. The abnormal concentration of AA can be related to various kinds of diseases including scurvy, mental disorder, cancer, cold, AIDS, and digestive disorder [15]. In addition, AA has been found to participate in a lot of life processes, such as cell division, iron absorption, acceleration of collagen synthesis, melanogenesis inhibition, and metabolism [69]. Since AA is so important, it is of great significance to develop a rapid, sensitive analytical method for the detection of AA. Up to now, several methods have been applied for AA detection, such as chemiluminescence, fluorescence, electrochemistry, chromatography, capillary electrophoresis and colorimetry [1015]. Of these, the electrochemical method has attracted much interest owing to its simple operation, fast response, high accuracy, high selectivity, and effectiveness.

Dopamine (DA) and uric acid (UA) coexist with AA in a biological environment [1618]. Since the oxidation peak potential of DA and UA is close to that of AA, they often interfere with the electrochemical detection of AA. To solve this problem, researchers have used a variety of methods, among which the electrochemical sensor method [1925] exhibits excellent anti-interference effect, which can effectively separate the peaks of AA, UA, and DA. However, there are few reports on the application of CQD modified electrode in the electrochemical detection of AA.

CQDs are a novel category of carbon nanomaterials [26] that have drawn considerable attention in recent years due to its excellent properties including low toxicity [27], nonpollution [28], high stability [29], simple preparation [30], and extensive carbon sources. Until now, CQDs have been used in the eletrochemical detection of dopamine [31], L-cysteine [32], curcumin [33], and antipsychotics [34]; the results of these studies demonstrated that CQDs have great application prospects in the eletrochemical field. To date, methods synthesizing the CQDs have been studied extensively, such as hydrothermal [35], laser ablation [36], ultrasonic [37], microwave synthesis [38], and solvothermal [39]. Among them, the microwave method is simple, short time-consuming, environmentally friendly, and low cost.

In this paper, we have developed a sensor for AA determination based on CQDs, which were synthesized using glucose as the carbon source and polyethylene glycol-200 as the surface passivator through a one-step facile microwave approach. This fabricated CQD electrode demonstrated wider linear ranges and higher sensitivity toward AA compared to other modified electrodes based on electrochemical strategies. At the same time, the as-prepared sensor has successfully separated the oxidation peak of AA, DA, and UA. The electrochemical response of the as-prepared sensor was greatly improved and finally realized the selective detection of AA. We envision that a CQD-based sensor will become a crucial platform for determination of AA in a complex system in the future.

2. Experimental

2.1. Reagents

AA was purchased from Lianxing Biotechnology (Tianjin), polyethylene glycol-200 was bought from Beijing Solarbio Science & Technology, dopamine hydrochloride was purchased from Sigma-Aldrich, UA was obtained from Ron Chemical Reagent Company (Tianjin), glucose was obtained from Tianjin Yingda Xigui Chemical Reagent Factory, and 0.1 M phosphate buffer solution (PBS) (pH 7.0) was prepared by mixing a proper proportion of Na2HPO4 and NaH2PO4. All the other reagents were prepared with ultrapure water and directly used without further treatment.

2.2. Apparatus

All electrochemical measurements were carried out at the AMETEK PARSTAT 4000 electrochemical workstation (AMETEK Commercial Enterprise (Shanghai) Co., Ltd. Beijing Branch). All electrochemical experiments were performed with a three-electrode cell with a modified GCE as the working electrode, a platinum counter electrode (CE), and a saturated calomel electrode (SCE) as the reference electrode. A magnetic stirrer was acquired from Ronghua Instrument Manufacturing (Jiangsu, China). Field emission scanning electron microscopy (FESEM) images were obtained from Nova NanoSEM 430 (FEI, USA). High-resolution electron microscopy (HRTEM) and energy-dispersive spectrometer (EDS) micrographs were acquired by using FEI Talos F200X equipped with an energy-dispersive spectrometer analyzer. CQDs were centrifuged using a LX-400 centrifuge (Kylin-Bell Lab Instruments Co., Ltd.), microwave reactor (LG Electronics, China), and dialysis bags (Beijing Solarbio Science & Technology). All DPV measurements in this paper were performed by applying a sweep potential from -1.5 V to +1.5 V at pulse width of 0.2 s and an amplitude of 50 mV.

2.3. Synthesis of CQDs

The CQDs were synthesized by one-step microwave heating according to literature [40] with a little modification. 20 ml of polyethylene glycol-200 was added into a mixture of 6 ml of ultrapure water and 4 g of glucose under stirring condition until the solution became transparent and colorless; then, the mixed solution was placed in a microwave reactor under medium heat condition for 3 minutes. The obtained dark viscous solution was cooled down to room temperature, and then, the appropriate amount of the solution was put into a dialysis bag (cutoff Mn: 100-500 Da) for 24 hours. Then, the obtained light brown solution was centrifuged at 6000 r/min for 15 min, bottom black precipitation was removed, and the upper solution was collected and stored in the refrigerator at 4°C.

2.4. Preparation of CQD Modified GCE (CQDs/GCE)

Before use, the GCE was polished to mirror with 0.05 μm alumina slurry, and then, the bare electrode was modified in CQD solution with a three-electrode system using CV under the conditions of -1~2 V, 20 cycles, at 50 mV/s. After modification, the electrode was washed with ultrapure water and dried for later use.

3. Results and Discussion

3.1. Characterization of CQDs

Morphology of CQDs was characterized by a high-resolution transmission electron microscope (HRTEM), field emission scanning electron microscope (FESEM), and energy-dispersive spectrometer (EDS). As is shown in Figure 1(a), the CQDs were found to have a quasispherical shape with a size lower than 10 nm, which is consistent with the results reported by previous literature [41, 42]. It can be seen from the inset of Figure 1(a) that the prepared CQDs possess a lattice structure with a lattice spacing of 0.14 nm, which can be attributed to the (102) diffraction plane of graphite carbon [43]. From the EDS analysis image (Figure 1(d)), the main element composition of the CQDs is C, and the peaks of Cu, O, and Si should be originated from the substrate. These results confirmed that the CQDs have been successfully synthesized. Moreover, after the CQDs were electropolymerized on the electrode surface (Figures 1(b) and 1(c)), the size of the CQDs increased (close to one hundred nanometers), which indicates that the electrochemical properties of CQDs can be improved by electropolymerization.

3.2. Electrochemical Characterization

The electrochemical properties of the bare GCE and CQDs/GCE were characterized by CV using 20 mM Fe(CN)63-/4- containing 0.1 M KCl as the electrochemical probe. As can be seen from Figure 2(a), compared with bare GCE (curve A), the peak current of CQDs/GCE (curve B) increased obviously and the interspike interval was decreased. It may be that the surface structure of CQDs can accelerate the electron transfer rate of the redox system and increase the adsorption of AA at the sensor interface. These results suggest that the prepared CQDs have good properties.

The EIS of different electrodes was measured under the conditions of 10-2~105 Hz, 0.24 V bias potential, and 5 V amplitude in 20 mM Fe(CN)63-/4- containing 0.1 M KCl. The direct part of the figure represents the diffusion process, and the diameter of the semicircle shows the resistance of the electron transfer. As shown in Figure 2(b), the CQDs/GCE had a smaller semicircle (curve B) than GCE (curve A), which suggested that the resistance of the modified electrode was significantly decreased, indicating that CQDs/GCE has better conductivity. The result was consistent with Figure 2(a).

DPV was also used to further evaluate the detection performance of the modified electrode in 0.1 M PBS (pH 7.0) containing 0.5 mM AA. Figure 2(c) showed that there was a broad but inapparent oxidation peak (curve A) on GCE and a narrow but obvious oxidation peak (curve B) on CQDs/GCE. The oxidation peak current of CQDs/GCE was about three times larger than that of the bare electrode, which indicated that CQDs/GCE had obvious electrocatalytic oxidation effect on AA.

3.3. Effect of Scan Rate of AA

In order to further study the reaction mechanism of AA on the as-prepared sensor surface, the effect of the scan rate on electrochemical behavior of AA was investigated. From Figure 3(a), it was observed that the catalytic current of AA on CQDs/GCE increased gradually with the increase of the scan rate, and the peak current of redox showed a good linear relationship with the scan rate in the range of 50-250 mV s-1 (Figure 3(b)), and the linear regression equation is demonstrated as the following, respectively: (), (). The results indicated that the reaction process of the modified electrode was controlled by adsorption. In addition, the oxidation peak potential and reduction peak potential almost did not change, indicating that CQDs/GCE had a good reversibility.

3.4. Electrochemical Sensing of AA on the CQDs/GCE

The electrochemical response of CQDs/GCE toward AA was studied by DPV. In the 0.1 M PBS solution, a certain amount of AA solution was added sequentially. As shown in Figure 4(a), at low concentration, the oxidation peak potential of AA remained at 0 V, and then, with the increase of AA concentration, the oxidation peak of AA existed positive transfer, which should be related to the change of the pH of the detection solution resulting from the accumulation of the oxidation product of AA. In addition, with the increase of AA concentration, the peak current of AA also increased. When AA concentration ranges from 0.01 mM to 3 mM () and from 4 mM to 12 mM (), the oxidation peak current has a good linear relationship with AA concentrations (Figure 4(b)), with a detection limit of 10 μM (). Compared to many previous works based on electrochemical strategies that have been reported (Table 1), the modified electrode exhibited wider linear range and higher sensitivity. The catalytic mechanism for AA oxidation is shown in Scheme 1. The redox reaction of AA on the CQDs/GCE sensor is an electrochemical process based on two-proton and two-electron transfer, accompanied by the formation of dehydroascorbic acid.

3.5. Anti-interference of the Sensor

The anti-interference ability of the sensor to AA in the presence of DA and UA was investigated by DPV. As shown in Figure 5(a), three well-separated peaks were obtained at CQDs/GCE (curve B) while bare GCE (curve A) cannot separate the peaks of AA and DA. To further prove that the CQDs/GCE has good selectivity to AA, we have set control groups that only change the concentrations of AA (Figure 5(d)), DA (Figure 5(b)), and UA (Figure 5(c)) while keeping two other substances constant and changing the concentrations of AA, DA, and UA simultaneously (Figure 5(e)). These results showed that the existence of DA and UA could not interfere with the electrochemical detection of AA on the CQDs/GCE electrode. In addition, several possible interfering substances in the real sample such as Na+, Cl-, Mg2+, SO42-, and glucose at 100-fold concentration were also examined by chronoamperometry technique at the applied potentials of +0 V, and different substances were added every 100 seconds (Figure 5(f)), which further indicated that the prepared sensor has good selectivity for the detection of AA.

The reusability of the as-prepared sensor was investigated by six modified electrodes (Figure 6(a)) that were fabricated under the same condition in 0.1 M PBS (pH 7.0) containing 0.5 mM AA. The relative standard deviation of peak current was calculated as 3.58%, which indicated that the sensor has good reproducibility. The sensitivity of the sensor was calculated as the slope line of each electrode area (0.2826 cm2), equal to 44.13 and 9.66 μA-1μM-1 cm-2, respectively. As shown in Figure 6(b), the long-term durability of CQDs/GCE was measured by chronoamperometry [5, 59] under the conditions of 2000 s and 0 V; the current of the sensor decays rapidly at the beginning, which may be due to the rapid adsorption of a large number of AA on the sensor surface that makes the number of active sites reduced quickly. After that, the adsorption of AA and the release of active sites basically reached a balance, so the current of the sensor tended to be stable [19]. Since the current decay only occurs in the first few seconds, and then remains stable for a long time, it proves that the sensor has good stability.

3.6. Real Sample Analysis

To access the practical application value of the CQDs/GCE sensor, DPV was used to detect the fetal bovine serum diluted 10 times with 0.1 M PBS. The experimental results are shown in Table 2. The recoveries of AA were in the range of 99.3%-102.6%, and the relative standard deviation was lower than 3%, indicating that the sensor has important practical application of significance.

4. Conclusions

The CQDs were successfully synthesized using a microwave reactor, and the sensor based on CQDs was successfully constructed. The CQDs/GCE had two wide linear ranges, excellent sensitivity, and good stability. In particular, its outstanding anti-interference ability further indicated that CQDs are promising candidates for AA detection in real sample analysis and are expected to be used for simultaneous detection of AA, DA, and UA.

Data Availability

The generated or analysed data used to support the findings of this study are included within the article.