Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements

Figure 2

Bland and Altman analysis: agreement between GE Healthcare Lunar Prodigy and iDXA scans. The difference in BMD values between two scans as a function of the mean value for the two DXA scanners at the (a) femoral neck, (b) total hip, (c) lumbar spine (L2–L4), and (d) total body. Mean difference () = straight line; limits of agreement () = dashed lines. ((a) to (c)); (d). The total body BMD difference was strongly dependent on mean BMD: at high BMD values iDXA showed significantly higher values than Prodigy, whereas at low BMD values the opposite was found. The hip and spine BMD difference was less dependent on the mean BMD values.
(a)
(b)
(c)
(d)