Research Article

Correlations among Corneal Biomechanical Parameters, Stiffness, and Thickness Measured Using Corvis ST and Pentacam in Patients with Ocular Hypertension

Table 4

Multiple line regression analysis associated factors for significant Corvis ST parameters.

Dependent variableIndependent variableStandardized regression coefficient

A1VPre-IOP−2.85<0.001
Changes in IOPs−3.33<0.001
Axial lengths−0.040.771
Corneal thickness in central pupil−0.350.021

A1TPre-IOP2.030.007
Changes in IOPs2.530.002
Axial lengths−0.130.449
Corneal thickness in central pupil0.320.062

A2TPre-IOP−1.990.020
Changes in IOPs−2.250.011
Axial lengths−0.280.152
Corneal thickness in central pupil0.160.400

Changes in AVPre-IOP0.0080.044
Changes in IOPs−0.0150.000
Axial lengths0.0110.072
Corneal thickness in central pupil−0.0010.003

Changes in ATPre-IOP−0.0170.690
Changes in IOPs0.2850.000
Axial lengths−0.1030.113
Corneal thickness in central pupil0.0020.608

PDHCPre-IOP−1.830.005
Changes in IOPs−2.220.001
Axial lengths0.370.015
Corneal thickness in central pupil0.050.736

DAHCPre-IOP−1.530.050
Changes in IOPs−1.980.015
Axial lengths0.230.216
Corneal thickness in central pupil−0.170.343

DA ratioPre-IOP−1.190.070
Changes in IOPs−1.510.026
Axial lengths−0.160.309
Corneal thickness in central pupil−0.80<0.001

SP-A1Pre-IOP1.390.030
Changes in IOPs1.880.005
Axial lengths0.070.648
Corneal thickness in central pupil0.74<0.001

Inter radiusPre-IOP−1.480.023
Changes in IOPs−1.900.005
Axial lengths0.020.904
Corneal thickness in central pupil−0.77<0.001

All the independent variable from liner regression analysis were included in the model. Pre-IOP: Pre-Goldman intraocular pressure before antiglaucomatous medications application. Changes in IOPs: changes in Goldman intraocular pressure after antiglaucomatous medications applications. A1V: applanation1 velocity; A1T: applanation1 time; A2T: applanation2 time; Changes in AV: changes in A1V-A2V after anti glaucomatous medications applications; Changes in AT: changes in A1T-A2T after antiglaucomatous medications applications; PDHC: highest concavity peak distance; DAHC: highest concavity deformation amplitude; SP-A1: stiffness parameter A1. Multiple liner regression analysis of different independent variable with .