Comparison of Two Wavefront Autorefractors: Binocular Open-Field versus Monocular Closed-Field

Table 5

Intersession repeatability of oblique cylindrical vector (J45) obtained with both binocular open-field (Eye Refract) and monocular closed-field (VX110) autorefractors in terms of mean difference between sessions (bias), its standard deviation (SD), repeatability (Sr), and its 95% limit (r).

Wavefront autorefractorGroupJ45Session 1-session 2Session 1–session 3Session 2-session 3Repeatability [95% limit] (D)

Eye refractTotalBias ± SD (D)−0.02 ± 0.20−0.01 ± 0.150.01 ± 0.110.11 [0.30]
value0.8121.0001.000
TeenagersBias ± SD (D)0.00 ± 0.060.02 ± 0.100.02 ± 0.100.06 [0.18]
value1.0001.0001.000
AdultsBias ± SD (D)0.00 ± 0.100.00 ± 0.090.00 ± 0.070.06 [0.18]
value1.0001.0001.000
PresbyopesBias ± SD (D)−0.05 ± 0.28−0.04 ± 0.200.01 ± 0.130.15 [0.41]
value0.6710.5651.000

VX110TotalBias ± SD (D)0.02 ± 0.100.02 ± 0.070.00 ± 0.100.06 [0.18]
value0.2150.0311.000
TeenagersBias ± SD (D)0.02 ± 0.070.01 ± 0.060.00 ± 0.070.04 [0.12]
value0.8310.7961.000
AdultsBias ± SD (D)−0.01 ± 0.070.00 ± 0.060.01 ± 0.070.04 [0.12]
value1.0001.0001.000
PresbyopesBias ± SD (D)0.04 ± 0.120.03 ± 0.08−0.01 ± 0.120.08 [0.77]
value0.1600.0381.000

The statistical comparison was done between sessions. , one-way ANOVA for paired samples with Bonferroni correction.