Comparison of Repeatability and Agreement between Swept-Source Optical Biometry and Dual-Scheimpflug Topography

Table 1

Repeatability of various biometric parameters with two biometry devices.

ParametersWithin-subject standard deviationCoefficient of variation (%)ICC (95% CI)
IOLMaster 700Galilei G6IOLMaster 700Galilei G6IOLMaster 700Galilei G6

AL (mm)0.0050.0370.0210.1471.000 (1.000-1.000)0.999 (0.998–0.999)
ACD (mm)0.0080.0180.2310.5680.999 (0.998–0.999)0.996 (0.993–0.997)
Kf (D)0.0980.2820.2280.6640.989 (0.984–0.992)0.956 (0.938–0.970)
Ks (D)0.0760.2020.1720.4580.997 (0.996–0.998)0.980 (0.971–0.986)
Km (D)0.0690.2100.1580.4860.996 (0.994–0.997)0.973 (0.962–0.982)
WTW (mm)0.1100.0910.8540.7680.879 (0.838–0.912)0.832 (0.769–0.882)
CCT (μm)1.8734.1480.3490.7310.995 (0.993–0.996)0.965 (0.950–0.976)
LT (mm)0.0110.040.3320.9830.999 (0.999–0.999)0.984 (0.977–0.990)

AL: axial length; ACD: anterior chamber depth; Kf: flat keratometry; Ks: steep keratometry; Km: mean keratometry; WTW: white-to-white corneal diameter; CCT: central corneal thickness; LT: lens thickness; ICC: intraclass correlation coefficient.