Abstract

Steroid inhalation is the standard bronchial asthma therapy and it includes powdered metered doses, dry powder, and nebulizer suspension. However, particle sizes vary widely. The research goal was to demonstrate that different budesonide administration forms and devices have various deposition rates in the airway obstruction region. Here, we compared relative inhalation therapy efficacies and identified therapies that delivered the highest drug doses to the airway obstruction region. Weibel’s anatomy data were used to identify the airway obstruction region in asthma. Based on European Standardization Committee data, we investigated the diameters of the drug particles being deposited there and evaluated the average particle size and distribution of the budesonide dosage forms and application devices. Drug dose depositions were measured by HPLC at each stage of a Cascade Impactor. Weibel’s anatomy data indicated that the 1st–4th bronchial generations comprised the airway obstruction region and corresponded to the tracheobronchial area. According to the European Standardization, particles 2–6 µm in diameter were readily deposited there. The proportions of particles in this size range were 33.0%, 32.0%, 59.0%, and 78.0% for Turbuhaler, Symbicort, mesh-type NE-U22 suspension, and jet-type NE-C28 suspension, respectively. We localized the airway obstruction regions of bronchial asthma and identified the optimal inhalation therapy particle size. An electric nebulizer was more efficacious for budesonide administration than dry powder delivery. The NE-C28 treatment deposited 2.36x more budesonide in the airway obstruction region than dry powder delivery systems.

1. Introduction

Steroid inhalation therapy is the most popular treatment for both pediatric and adult bronchial asthma. Steroid inhalation is the most effective currently available therapy for bronchial asthma. It has other benefits such as inhibition of the development of serum-specific IgE [1] in childhood asthma. There are three main types of bronchial asthma inhalation methods: powdered metered dose inhalation (PMDI), dry powder inhalation (DPI), and electronic nebulizer suspensions (jet, ultrasonic, and mesh types). We propose that electronic nebulizer inhalation is the most effective of these three methods [2].

Bronchial asthma is obstructive and reacts to β2-stimulator disease without lung remodeling. To improve steroid inhalation efficacy, it is necessary to increase the amounts of steroids reaching the asthmatic obstruction regions. Certain factors such as steroid particle size (MMAD; mass median aerodynamic diameter), particle density, particle dosage form, particle surface tension, temperature, atmosphere, and airflow velocity affect steroid dose deposition. Among these factors, particle size is the most important because it significantly influences the deposition region.

Budesonide is marketed and supplied in two dosage forms. It is, therefore, possible to compare the two. The site and amount of inhaled antiasthma drug deposition are important in evaluating asthma treatment efficacy. However, few studies used the same setups and conditions with different drug forms or devices. Therefore, in this study, the amounts of budesonide deposited in each stage of the Marple-type Cascade Impactor were evaluated by delivering the drug as a DPI or an inhalant suspension (jet-type NE-C28 or mesh-type NE-U22 nebulizer). Particle size distributions and bronchial deposition rates were evaluated. This enabled the comparison of the relative efficacies of the various therapies and further identified the therapy that delivered the highest drug doses to the airway obstruction region.

2. Materials and Methods

2.1. Weibel’s Anatomy Data

Weibel anatomy data [3] (Table 1), which show the relationships between total cross-sections (Figure 1) and bronchial generations in bronchial asthma, were used to identify the airway obstruction regions. Weibel’s respiratory tree model was first reported in 1963 and adopted in the ICRP respiratory tract textbook [4]. It is still cited in papers published in the twenty-first century.

Obstructed lung regions were determined from the Weibel anatomy data. In this model, the airway repeatedly bifurcates. The airways from the 1st to 4th generations are called bronchi.

Those from the 5th to 16th generations are known as bronchioles. The airways from the 17th to 19th generations are referred to as respiratory bronchioles, while those from the 20th to 23rd generations are the alveolar ducts and sacs. The luminal walls of the 17th and more distal generations consist of the alveolar wall. Therefore, the airways from the 1st to 16th generations are known as bronchioles. Those from the 8th to 16th generations which are <2 mm in diameter are known as small airways or peripheral bronchi. These initially bifurcate by reducing their internal diameter at a rate of 1/√2 but, the tapering rate gradually decreases in more distal areas. Therefore, total cross-sectional area increases with airway distality. Compared with the tracheal cross-sectional area (2.54 cm2), those of the 4th, 8th, 11th, and 16th generations are 0.98x, 2.7x, 11.3x, and 70.8x in size, respectively.

2.2. Total Cross Section and Obstructive Region

Airway diameters were calculated using one duct from the total Weibel’s cross-section (Figure 1). Figure 1 shows that the diameters of the 1st–4th bronchial generations are mainly small and constitute a physiological airway obstruction region. The airways of the 1st–4th generations’ obstruction sites serially bifurcate and are the narrowest according to the calculations of one duct from the total cross-sectional area. It was, therefore, considered that the asthmatic airway in Figure 1 showed the diameter of one duct based on the total cross-section obstruction region when it was localized to the 1st–4th bronchial generations. It is obvious from Figure 1, that the airway of 1st–4th generations is the most narrow of the total airways in the lung. Eosinophilic inflammation [5] occurred in all lung airways, but was limited in the airway obstruction region.

2.3. Aerosol Particle Size Deposits on Obstructive Region and European Standardization Committee

The sizes of the particles deposited in the airway obstruction regions (1st–4th bronchial generations) are determined from the European Standardization Committee (CES) (Table 2) and the Cascade Impactor (Figure 2) data. The aerosol particle sizes and the pulmonary deposition areas were obtained from the European Standardization Committee data (CES) (EN 13544-1) [6] (Table 2). The Aerosol particle sizes were defined in terms of mass median aerodynamic diameter (MMAD) and geometric standard deviation (GSD). The consensus was that the efficacious aerosol particle diameter for asthmatic inhalation treatment is in the range of 2–6 µm (Table 2).

2.4. Andersen Instrument (Cascade Impactor Manual Data)

The regions of trachea and primary bronchi are almost matched with 1st–4th bronchial generations. On trachea and primary bronchi, particles of 3.3–4.7 µm are deposited. The MMAD on Andersen Instrument is 4.0 µm as normal distribution, which quite matched with the MMAD of the European Standardization Committee. In the next step, we chose the particle distribution of CES, because the data of CES are supported by thousands of published papers.

2.5. Particle Size Analyzed by Cascade Impactor

The Cascade Impactor (Andersen Instruments, Inc.) is used mainly to analyze the atmosphere, [7] but it is also now used to simulate the lung airway (Figure 2). The Marple Personal Cascade Impactor 298 [8, 9] was used (Figures 3(a)3(c)). Its specifications are as follows: cutoffs at impactor stages 1–7, and 8: 21.3, 14.8, 9.8, 6.0, 3.5, 1.55, 0.93, and 0.52 µm, respectively, flow rate: 2 L/min (nominal), 0.5–5 L/min (lower and upper limits), and 1–3 L/min (recommended). The device is constructed from precision-machined aluminum and the impactor stages are nickel plated. Stages 7 and 8 have holes rather than slits to simulate real bronchial bifurcations. At the more distal sites of the airway model, the total cross-sectional area of the slits decreases and the flow velocity increases. For the bronchial bifurcations, however, the total cross-sectional area increases at the more distal sites, reaching a maximum at the terminal alveolar generation where the flow velocity falls to zero.

2.6. High-Performance Liquid Chromatography

For each Cascade Impactor stage, the budesonide trapped on the GF/A filters was assayed by HPLC (Shimadzu LC-10ADVP; Shimadzu Corp., Tokyo Japan) under the following conditions.(1)Column: Shim-pack VP-ODS 150 mm × 4.6 mm.(2)Mobile phase: Methanol : acetic acid 0.1% (7 : 3).(3)Retention time: 8 min.

The airway deposition rate was evaluated according to a particle size spectrum using the Cascade Impactor.

2.7. MMAD and Particle Size Distribution of Dry Powder and Suspension of Budesonide

The tracheobronchial regions correspond to the size of 1st–4th bronchial generations. Therefore, from CES, particles 2–6 µm in diameter are expected to deposit in the airway obstruction regions. The mass median aerodynamic diameters (MMAD) of the particle size distribution were 2.20 µm for the Turbuhaler (DPI) (Figure 4(a)), 2.60 µm for the Symbicort (DPI) (Figure 4(b)), 4.52 µm for the mesh-type NE-U22 suspension inhalant (Figure 4(c)), and 3.84 µm for the jet-type NE-C28 suspension inhalant (Figure 4(d)) (Table 1).

The experimental conditions were applied based on the European standard (EN13544-1 2007) Annex CC. According to EN 1354-1(2007), the ambient condition shall be 23 ± 2, Relative humidity 45–75% Rh. Thus the data were obtained at 22, 50% Rh.

3. Discussion

3.1. Deposition Rate on Obstructive Region by Dry Powder and Suspension of Budesonide

All these measurements were within the 2–6 µm range. Therefore, the particles should have been deposited in the bronchi. All four distribution patterns were unimodal. The percentages of particles within the 2–6 µm diameter range were 33.0% for the Turbuhaler, 32.0% for the Symbicort, 59.0% for the mesh-type NE-U22 suspension, and 78.0% for the jet-type NE-C28 suspension (Table 1). The percentages of particles generated by the electric nebulizers within the 2–6 µm diameter range were higher than those generated by the DPIs. Differences in budesonide deposition rate and distribution were compared for suspension- (jet and mesh nebulizers) and dry powder (Pulmicort and Symbicort) inhalation methods using the Cascade Impactor. The data derived from the Cascade Impactor describe the relationship between aerosol particle size and bronchial generation. The 4.7–7.0, 3.3–4.7, 2.1–3.3, 1.1–2.1, and 0.6–1.1 deposition sites correspond to the pharynx, trachea, and primary bronchi, secondary bronchi, terminal bronchi, and alveoli, respectively. The bronchial obstruction site consisting of the 1st to 4th generations according to Weibel’s anatomy data nearly aligns with the trachea and the primary obstruction site (deposition diameter range: 3.3–4.7 µm). In these studies, the data from the two sources were compared. The 2–6 µm diameter range was taken as the optimal deposited particle size in the airway obstruction region in bronchial asthma because it is corroborated by many literature citations.

3.2. Particle Size in Current pMDI

The pMDIs react with CFCs and form CFC-pMDIs which are then converted to HFA-pMDIs [10]. HFA + alcohol-pMDI is a super-small aerosol which can effectively reach and treat small airway obstructions. The MMADs of beclomethasone pMDI (Qvar, Sumitomo Dainippon Pharma Co., Ltd., Tokyo, Japan) and ciclesonide pMDI (Alvesco, Teijin Pharma, Tokyo, Japan) are 1.1 µm and 0.9 µm by each statement of virtues of a medicine, respectively.

Certain pulmonologists recommend that super-small steroid particles are efficacious in the management of small airway (bronchial diameter <2 mm) obstruction [11, 12]. However, the total cross sectional areas of the 8th (Weibel anatomy data; (bronchial diameter 1.86 mm; small airway; calculated from the total cross section as a single duct)), 9th (1.54 mm), 10th (1.3 mm), and 11th (1.09 mm) bronchial generations were 6.95 cm2, 9.65 cm2, 13.4 cm2, and 19.6 cm2, respectively, which are far larger than the airway obstruction regions (<2.5 cm2). Therefore, small airway obstruction may, in fact, be absent in bronchial asthma, and the particle diameters of beclomethasone- and ciclesonide pMDI are too small to enable these drug forms to be efficacious for the treatment of this condition. In this study, the most major defect point affected is only inspiratory, not expiratory phase by the Cascade Impactor. About expiratory phase, no examinations could be performed. Bronchial asthma is an expiratory distress disease, mechanism of expiratory distress must be examined. According to those examinations performed, the identity of super small aerosol released pMDI may be useful to small airway obstruction.

3.3. Estimation of Inhaled Steroid Efficacy by One Factor as Particle Size is Difficult

In theory, the cutoff diameters of the particles deposited by inertial collision are determined at each stage. Particles 2–6 µm in diameter are deposited in the bronchi/bronchioles. Particles are deposited on a Cascade Impactor primarily by inertia, whereas deposition which actually occurs by diffusion or sedimentation in the terminal airway of a living lung is not considered. Consequently, an airway tree model which more nearly approximates living lungs is required.

The percentage of particles <2 µm in diameter deposited in the alveoli was 46.0% for the Turbuhaler, 42.0% for the Symbicort, 11.0% for the mesh-type NE-U22 suspension, and 17% for the jet-type NE-C28 suspension. The percentages of particles <2 µm in diameter generated by the DPI (Pulmicort and Symbicort) were higher than those generated by the mesh- and jet-type electric nebulizers.

Evaluation of the efficacy of steroid inhalation for asthma treatment is very difficult. Relative to the cromoglycate inhalation therapy, the Pulmicort suspension has greater clinical efficacy [13].

However, certain reports [1416] describe differences in clinical efficacy between Symbicort and ADVAIR. Nevertheless, these assessments are challenging because clinical symptom scores are unstable and the clinical visit continuity is very low [17]. To this end, we devised a new asthma inhalation therapy effectiveness evaluation system based on aerosol particle size and distribution. At any rate, the final evaluation must nonetheless be based on the degree of clinical improvement.

3.4. On Budesonide Inhalation Therapy, Nebulizer Inhalation Treatment by Pulmicort Suspension is the Best Method (Table 3)

The Cascade Impactor was previously used to assess cromoglycate efficacy (DPI, pMDI, and suspension) [18]. In the present study, the effectiveness of budesonide (DPI and suspension) was evaluated. The mass median aerodynamic diameter (MMAD) of particle size distribution was 2.20 µm for the Turbuhaler (DPI), 2.60 µm for the Symbicort (DPI), 4.52 µm for the mesh-type NE-U22 inhalant, and 3.84 µm for the jet-type NE-C28 inhalant (Figure 3). Particles within the 2–6 µm diameterrange were deposited in the bronchi. The distribution pattern was unimodal for the other dosage forms. The percentages of particles 2–6 µm in diameter generated by the Turbuhaler, Symbicort, mesh-type NE-U22 suspension, and jet-type NE-C28 suspension were 33.0%, 32.0%, 59.0%, and 78.0%, respectively (Table 2). The electric jet nebulizer delivered ~2.36x more drug to the bronchi/bronchioles than the Turbuhaler. The normal budesonide dose delivered with the Pulmicort 200 Turbuhaler is 400 µg/d; whereas, with the Pulmicort Suspension it is 1,000 µg/d. Based on deposition rates of 33% with the Turbuhaler and 78% with the NE-C28 jet nebulizer suspension, the rates of budesonide dose deposition to the airway obstruction area are 132 µg/d and 780 µg/d, respectively. In fact, the rate of budesonide deposition by NE-C28 into the bronchial obstruction area is 5.9x higher than that delivered by the Turbuhaler. The results of the present study suggest that NE-C28 inhalation is the most efficacious method of delivering budesonide for asthma therapy.

Differences in the diameters of the inhaled budesonide particles result in variable degrees of improvement in the clinical symptoms of bronchial asthma. We investigated reports on comparative variations in bronchial asthma mitigation in response to the inhalation of the dry powder and suspension forms of budesonide. Nevertheless, no studies were found comparing clinical improvements using different budesonide dosage forms and application devices. When equivalent budesonide doses in DPI and suspension form (NE-C28) are administered, the latter has an obvious advantage in terms of clinical bronchial asthma improvement because it deposits 2.36x more budesonide than the former. Budesonide inhalation routes, application devices, and doses merit further research.

Currently inhaled asthma drugs have a particle diameter range of 0.9–10 µm. Compared the volume or gravity of big and small aerosols in current inhaled drugs, the big aerosol of volume or gravity is reached to 1000 times as cubic of diameter scale. A future study objective is to optimize the amount of inhalation drug which reaches the airway obstruction region including small airway (16, 17) in bronchial asthma.

4. Conclusion

Among the various dosage forms of budesonide, the NE-C28 suspension may deposit ≤2.36x more drug in the bronchial obstruction region than the Pulmicort Turbuhaler (DPI). The NE-C28 suspension inhalation therapy, then, may be the most efficacious type of budesonide inhalation therapy for bronchial asthma. The particle size (MMAD: 3.84 µm) and distribution (2–6 µm: 78%) obtained for the jet-type nebulizer suspension may have the greatest access to the airway obstruction region in bronchial asthma. The particle size (MMAD: 2.20) and distribution (<2 µm: 46%) obtained using the Turbuhaler (DPI) may be best suited for the treatment of peripheral lung diseases such as eosinophilic pneumonia, and COPD (chronic 260 obstructive pulmonary disease).

Abbreviations

CFC:Chlorofluorocarbon
COPD:Chronic obstructive pulmonary disease
DPI:Dry powder inhalation
HFA:Hydrofluoroalkane
MMAD:Mass median aerodynamic diameter
pMDI:Pressurized metered dose inhaler.

Data Availability

The data used to support the findings of this study are available from the corresponding author upon request.

Conflicts of Interest

The authors declare that they have no conflicts of interest and received no funding for this project. The cost of Cascade Impactor experimentation was covered by Omuron Health Care Corporation, Kyoto, Japan.

Supplementary Materials

In the present study, we compared different budesonide inhalation therapy dosage forms and application devices. First to fourth generation tracheal bronchi; particle size deposited: 2–6 µm; deposition rate: 78% (NE-C28 suspension). (Supplementary Material)