Research Article

Comparison of DMA-80 and ICP-MS Combined with Closed-Vessel Microwave Digestion for the Determination of Mercury in Coal

Table 4

Calibration curves and method detection limit (MDL) of Hg.

MethodsLinearity (ug/L)Determination coefficientMDL (ug/L)RSD (%)

50 mg sample (HNO3 + HF)0–1000.9999691.2200.34
50 mg sample (HNO3+HF + Au)0–1000.9999261.1070.29
50 mg sample (HNO3+HF + Au (after))0–1000.9999290.3100.94
100 mg sample (HNO3+HF + HCl)0–1000.9999291.0000.10