Research Article

Electron Density from Balmer Series Hydrogen Lines and Ionization Temperatures in Inductively Coupled Argon Plasma Supplied by Aerosol and Volatile Species

Table 2

Electron densities (/1015 cm−3) with their standard deviations ().

MatrixPNPN + CVGCVG
1000 W1200 W1000 W1200 W1000 W1200 W

HCl0.69 ± 0.031.05 ± 0.020.72 ± 0.011.09 ± 0.031.08 ± 0.021.27 ± 0.04
CH3COOH0.56 ± 0.021.07 ± 0.030.63 ± 0.041.10 ± 0.011.09 ± 0.031.32 ± 0.06