Research Article

New Kinetic Spectrophotometric Method for Determination of Fexofenadine Hydrochloride in Pharmaceutical Formulations

Table 3

Accuracy and precision for the determination of FEX in bulk powder by the proposed initial rate and fixed time methods.

MethodFEX, g mL−1RSD% Confidence limit 
(P = 0.05; n = 6)
Recovery%
TakenFoundaSD

Initial rate5.005.010.122.395.01 ± 0.13100.20
20.0020.200.200.9920.20 ± 0.22101.00
35.0034.900.280.8034.90 ± 0.3199.71
50.0050.130.140.2850.13 ± 0.15100.26

Fixed time5.005.040.132.585.04 ± 0.15100.80
20.0020.140.301.4920.14 ± 0.34100.70
35.0035.000.320.9135.00 ± 0.36100.00
50.0050.310.360.7150.31 ± 0.40100.62

Average of six determinations.