Research Article
New Kinetic Spectrophotometric Method for Determination of Fexofenadine Hydrochloride in Pharmaceutical Formulations
Table 2
Analytical characteristics of the fixed time (15 min) method.
| Parameters | FEX |
| (nm) | 610 | Beer’s law range (g mL−1) | 2.5–50.0 | Ringbom optimum concentration range (g mL−1) | 10.0–40.0 | Detection limit (g mL−1) | 0.055 | Quantification limit (g mL−1) | 0.183 | Molar absorptivity (L mol−1 cm−1) | 1.22 × 104 | Stoichiometric relationship, FEX : KMnO4 | 1 : 1 | Sandell’s sensitivity (g cm−2 per 0.001 absorbance unit) | 0.088 | Regression equationa | A = 0.0209C + 0.024 | Correlation coefficient, | 0.9998 |
|
|
A = mC + b, where A is the absorbance and C is the concentration in g mL−1.
|