The Accuracy and Precision of the Continuously Stored Data from Flash Glucose Monitoring System in Type 2 Diabetes Patients during Standard Meal Tolerance Test

Table 1

MARD in different categories.

MARD (%)MARD ≤ 30MARD ≤ 25MARD ≤ 20MARD ≤ 15MARD ≤ 10

Total (n = 208)10.7 ± 7.898.1%93.3%88.5%73.6%53.4%
Week 1 vs week 2
 Week 1 (n = 104)10.5 ± 7.399.0%94.2%90.4%76.9%52.9%
 Week 2 (n = 104)11.0 ± 8.397.1%92.3%86.5%70.2%53.8%
0.6610.3130.5800.3850.2710.889
VBG levels (mmol/L)
 3.9–10 (n = 76)11.2 ± 7.597.4%93.4%90.8%73.7%50.0%
 >10 (n = 132)10.4 ± 8.098.5%93.2%87.1%73.5%55.3%
0.4950.2700.9470.4250.9750.460
VBG times
 BG 0′ (n = 52)12.0 ± 8.296.2%92.3%86.5%67.3%44.2%
 BG 30′ (n = 52)11.2 ± 8.298.1%90.4%86.5%73.1%55.8%
 BG 60′ (n = 52)10.2 ± 7.598.1%94.2%88.5%78.8%57.7%
 BG 120′ (n = 52)9.5 ± 7.2100.0%96.2%92.3%75.0%55.8%
0.3550.9030.7920.7700.6030.498