Clinical Evaluation of the Peroral Cholangioscopy Using a New Videoscope

Received11 Sep 1998
Revised05 Jan 1999
Accepted05 Mar 1999

Abstract

Peroral cholangioscopy (PCS) has been performed in 22 cases using XCHF-B200 (Olympus Optical Co.) since June 1995 and in 77 cases using CHF-B20 (Olympus Optical Co.) after EST from Jan. 1989. XCHF-B200 has a longer rigid portion of distal end and a smaller channel diameter than CHF-B20. The successful rate of PCS using XCHF-B200 (82%) was lower than that of CHF-B20 (89%). The vascular pattern and fine vertical groove of the bile duct mucosa were shown more clearly on the photographs obtained with XCHF-B200 than those obtained with CHF-B20. However, not enough biopsy specimens could be obtained because the channel diameter of XCHF-B200 was rather small.If the length of rigid portion and biopsy channel of XCHF-B200 are improved, PCS using XCHF-B200 will be more useful for the diagnosis of bile duct disorders.

Copyright © 1999 Hindawi Publishing Corporation. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.