Research Article

Defining the Optimal Time of Adaptive Replanning in Prostate Cancer Patients with Weight Change during Volumetric Arc Radiotherapy: A Dosimetric and Mathematical Analysis Using the Gamma Index

Table 1

The changes in dosimetric parameters for (a) planning target volume coverage and (b) conformity and homogeneity indices with changes in body contour from −2.0 cm to 2.0 cm.
(a) Planning target volume coverage (%)

Body contour changes (cm)
−2.0−1.5−1.0−0.500.51.01.52.0

6-MV103.08102.33101.49100.66100.0099.2898.3197.3596.34
110.74108.94107.34106.06105.36104.72104.06103.52103.00
10-MV102.83101.83101.13100.51100.0099.4898.8198.0497.39
108.78107.50106.32105.48105.02104.68104.06103.56103.12
15-MV102.22101.66101.05100.46100.0099.5298.9398.3097.68
108.72107.56106.50105.62105.12104.60104.02103.54103.12

: the percentage of the prescription dose covering of the volume. Italic font: unacceptable.
(b) Conformity and homogeneity indices

Body contour changes (cm)
−2.0−1.5−1.0−0.500.51.01.52.0

6-MVCI1.181.141.101.051.000.910.620.330.19
HI1.0471.0371.0321.0291.0281.0291.0331.0371.043
10-MVCI1.151.111.081.031.000.930.740.500.31
HI1.0401.0391.0291.0281.0271.0281.0301.0331.036
15-MVCI1.141.111.071.020.990.940.780.550.39
HI1.0421.0371.0331.0311.0301.0261.0311.0331.034

CI, conformity index; HI, homogeneity index.