Research Article

Low-Dose and Scatter-Free Cone-Beam CT Imaging Using a Stationary Beam Blocker in a Single Scan: Phantom Studies

Table 1

Comparison of the averaged CT numbers and contrasts inside the contrast rods of the Catphan 600 phantom. The CT number errors are also shown in parentheses. The numbers of the ROIs are marked in Figure 3(d). All unites are in HU.

ROI1234567RMSE

Fan-beam CT−131−84−895248776−889−211
CBCT with correction−148−107−933227780−921−234
CT number (HU)(−17)(−23)(−38)(−21)(4)(−32)(−23)24
CBCT without correction−239−217−646−48202−639−281
(−108)(−133)(249)(−296)(−574)(250)(−70)250
CT number improvement9111021127557021847

CBCT with correction169118934201753933253
Contrast (HU)CBCT without correction8855452112322456126
Contrast improvement1.932.132.071.792.342.052.02