Personalization of CM Injection Protocols in Coronary Computed Tomographic Angiography (People CT Trial)
Table 3
Resulting injection parameters for the different groups.
Group BW (n = 112)
Group LBW (n = 108)
Group CO (n = 107)
value BW vs. LBW
value BW vs. CO
value LBW vs. CO
CM volume (mL)
47.5 ± 17.8 (18.4–100.2)
44.7 ± 13.9 (26.2–105.9)
42.7 ± 11.5 (27.0–101.3)
0.190
0.019
0.271
Test bolus (mL)
11.2 ± 4.0 (4.1–23.0)
10.3 ± 2.8 (6.0–20.9)
9.8 ± 2.2 (6.8–17.0)
0.047
0.002
0.192
Flow rate (mL/s)
5.1 ± 1.7 (2.1–9.1)
4.8 ± 1.2 (2.9–8.5)
4.6 ± 1.0 (3.2–9.1)
0.097
0.010
0.280
IDR (gI/s)
1.5 ± 0.5 (0.6–2.7)
1.4 ± 0.4 (0.9–2.6)
1.4 ± 0.3 (1.0–2.7)
0.085
0.008
0.306
TIL (gI)
14.3 ± 5.3 (5.5–30.1)
13.4 ± 4.2 (7.9–31.8)
12.8 ± 3.5 (8.1–30.4)
0.192
0.020
0.273
Peak pressure (psi)
111.5 ± 48.1 (30.0–262.0)
105.2 ± 39.6 (41.0–248.0)
98.9 ± 34.3 (52.0–275.0)
0.290
0.026
0.218
Note. Values are presented as mean ± standard deviation and ranges or numbers (percentages). BW = body weight; LBW = lean body weight; CO = cardiac output; CM = contrast media; IDR = iodine delivery rate; TIL = total iodine load. A significant difference was found between the group BW and group CO.