Using Residual Blood from the Arterial Blood Gas Test to Perform Therapeutic Drug Monitoring of Vancomycin: An Example of Good Clinical Practice Moving towards a Sustainable Intensive Care Unit

Table 1

Interday and intraday accuracy and precision of QC samples of vancomycin from LiHep samples.

SampleInterdayIntraday
Accuracy (recovery, %)Precision (RSD, %)Accuracy (recovery, %)Precision (RSD, %)

QC 12.2 mg/L100.41.897.92.5
QC 30.8 mg/L100.41.7100.01.5
QC 64.3 mg/L98.32.499.82.6

QC, quality control; LiHep, lithium heparin; RSD, relative standard deviation.