Research Article

New Mydriasis-Free Electroretinogram Recorded with Skin Electrodes in Healthy Subjects

Table 3

Normative values of ERG parameters.

ERG parametersImplicit time (ms)Amplitude (μV)
MeanRange (95% CI)MeanRange (95% CI)

Cone-responsea-wave12.112.0–12.25.85.6–6.0
b-wave28.228.1–28.321.120.4–21.7
Flicker-response 28.3 Hz24.624.6–24.721.621.0–22.2
Rod-responseScotopic b-wave96.595.4–97.744.443.2–45.6
Maximal-responsea-wave14.814.7–14.940.539.3–41.8
b-wave46.746.3–47.268.766.7–70.6
Oscillatory potentials152.5151.8–153.249.047.3–50.6

CI: confidence interval.