TY - JOUR A2 - Hu, Xiaoling AU - Ma, Heather Ting PY - 2014 DA - 2014/08/17 TI - A Blood Pressure Monitoring Method for Stroke Management SP - 571623 VL - 2014 AB - Blood pressure is one important risk factor for stroke prognosis. Therefore, continuous monitoring of blood pressure is crucial for preventing and predicting stroke. However, current blood pressure devices are mainly air-cuff based, which only can provide measurements intermittently. This study proposed a new blood pressure estimation method based on the pulse transit time to realize continuous monitoring. The proposed method integrated a linear model with a compensation algorithm. A calibration method was further developed to guarantee that the model was personalized for individuals. Variation and variability of pulse transit time were introduced to construct the compensation algorithm in the model. The proposed method was validated by the data collected from 30 healthy subjects, aged from 23 to 25 years old. By comparing the estimated value to the measurement from an oscillometry, the result showed that the mean error of the estimated blood pressure was -0.2±2.4 mmHg and 0.5±3.9 mmHg for systolic and diastolic blood pressure, respectively. In addition, the estimation performance of the proposed model is better than the linear model, especially for the diastolic blood pressure. The results indicate that the proposed method has promising potential to realize continuous blood pressure measurement. SN - 2314-6133 UR - https://doi.org/10.1155/2014/571623 DO - 10.1155/2014/571623 JF - BioMed Research International PB - Hindawi Publishing Corporation KW - ER -