Research Article

Evaluation of the MAScIR SARS-CoV-2 M Kit 2.0 on the SARS-CoV-2 Infection

Table 1

Repeatability and intermediate precision tests results of MAScIR SARS-CoV-2 kit M 2.0.

TargetNumber of values (N)Mean of CtSDCV%CV%Conclusion
Supplier notice

Repeatability
 Concentration of samples
  High concentrationRdRp514.520.392.691.56Compliant
S16.020.875.441.22Limit
CI21.940.331.491.69Compliant
  Medium concentrationRdRp527.610.632.291.70Compliant
S28.690.501.761.86Compliant
CI25.150.451.772.48Compliant
  Low concentrationRdRp533.240.732.201.70Compliant
S34.631.063.061.86Compliant
CI24.190.903.732.48Compliant

Intermediate precision
 Thermal cycler
  QS5RdRp2619.561.115.672.54Compliant
S20.641.095.271.87Compliant
CI21.951.185.393.49Compliant
  ExicyclerRdRp1321.620.753.462.54Compliant
S21.040.874.111.87Compliant
CI21.050.984.673.49Compliant
  Global intermediate precision (QS5 + Exicycler)RdRp3920.251.406.902.54Limit
S20.781.034.931.87Compliant
CI21.651.195.493.49Compliant

CV, coefficient of variation; SD, standard deviation.